首页> 外文期刊>Radiation Physics and Chemistry >THE RELATIVE RESPONSE OF FRICKE, DICHROMATE AND ALANINE DOSIMETERS TO ~(60)Co AND HIGH ENERGY ELECTRON BEAM RADIATION
【24h】

THE RELATIVE RESPONSE OF FRICKE, DICHROMATE AND ALANINE DOSIMETERS TO ~(60)Co AND HIGH ENERGY ELECTRON BEAM RADIATION

机译:弗里克,重铬酸盐和丙氨酸剂量对〜(60)Co和高能电子束辐射的相对响应

获取原文
获取原文并翻译 | 示例
           

摘要

The relative response of Fricke, dichromate and alanine dosimeters to ~(60)Co and high energy electron beam radiation has been determined directly using graphite calorimeters. The response of Fricke and alanine dosimeters is not significantly different in ~(60)Co or high energy electron beams. The response of dichromate dosimeters is, however, approximately 3% lower in electron beams than in ~(60)Co. Differences between the effective density of the dosimeter and the material surrounding it have been shown to be a significant source of error in electron beam dosimetry.
机译:Fricke,重铬酸盐和丙氨酸剂量计对〜(60)Co和高能电子束辐射的相对响应已使用石墨量热计直接确定。在〜(60)Co或高能电子束中,Fricke和丙氨酸剂量计的响应没有显着差异。然而,重铬酸盐剂量计的响应在电子束中比在〜(60)Co中低约3%。剂量计的有效密度与其周围材料之间的差异已被证明是电子束剂量测定中的重要误差源。