首页> 外文期刊>Oral Radiology >Spatial resolution of FineCube, a newly developed cone-beam computed tomography system
【24h】

Spatial resolution of FineCube, a newly developed cone-beam computed tomography system

机译:新开发的锥束计算机断层扫描系统FineCube的空间分辨率

获取原文
获取原文并翻译 | 示例
       

摘要

A new cone-beam computed tomography (CBCT) system called FineCube has been developed by Yoshida Dental Mfg. Co., Ltd. This CBCT system for dental use has a unique mechanism with a sliding U-arm, which enables a choice between a higher-resolution image and a wider-area image. This study was performed to evaluate the spatial resolution of the two scanning modes. First, a modulation transfer function (MTF) analysis was performed using the oversampling method. The results indicated that the high-resolution mode showed significantly higher spatial resolution than the wide area mode. Second, the resulting images were assessed after scanning a head computed tomography phantom using the two scanning methods, and profile analysis was performed on the images thus obtained, which confirmed the results of the MTF analysis. The results demonstrated that a shift in the rotation axis of only 146 mm yielded higher spatial resolution.
机译:吉田牙科株式会社开发了一种称为FineCube的新型锥束计算机断层扫描(CBCT)系统。这种用于牙科的CBCT系统具有独特的机制,带有滑动U形臂,可以在更高的范围内进行选择。分辨率图像和广域图像。进行这项研究以评估两种扫描模式的空间分辨率。首先,使用过采样方法进行了调制传递函数(MTF)分析。结果表明,高分辨率模式显示出比广域模式显着更高的空间分辨率。其次,在使用两种扫描方法扫描头部计算机断层摄影体模之后评估所得图像,并对由此获得的图像进行轮廓分析,这证实了MTF分析的结果。结果表明,仅146 mm的旋转轴偏移会产生较高的空间分辨率。