Erratum to 'The high-resolution time-of-flight spectrometer TOFTOF' [Nucl. Instr. and Meth. A 580 (2007) 1414-1422]

机译:勘误至“高分辨率飞行时间光谱仪TOFTOF” [Nucl。仪器和方法。 A 580(2007)1414-1422]

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The authors would like to correct the expression presented concerning the estimation of the energy resolution of TOFTOF spectrometer. The correct expression (4) is as follows: △E = (h~3)/(m_n~2λ_f~3) 1/(L_(sd)L_(pm)) (A~2 + B~2 + C~2)~(1/2), A = t_m(L_(pm) + L_(ms) + (λ_f~3)/(λ_i~3) L_(sd)), B = t_p(L_(ms) = (λ_f~3)/(λ_i~3) L_(sd)), C = (m_n)/h L_(pm)L_dλ_f, (4) where h is the Planck's constant, m_n the neutron mass, λ_i and λ_f are the wavelengths of the incident and scattered neutrons, respectively, and where the other parameters are given in Table 1. The authors would also like to acknowledge the prior work of Dr. Ana Gaspar, concerning the TOFTOF spectrometer and the estimation of its intensity and resolution functions [14], which is added hereby to the list of cited references. This citation refers to the introduction, the discussion of the instrument frame-overlap ratio (including in the caption of Fig. 6) and the discussion of the instrument resolution.
机译:作者想纠正提出的有关TOFTOF光谱仪能量分辨率估算的表达式。正确的表达式(4)如下:△E =(h〜3)/(m_n〜2λ_f〜3)1 /(L_(sd)L_(pm))(A〜2 + B〜2 + C〜2 )〜(1/2),A = t_m(L_(pm)+ L_(ms)+(λ_f〜3)/(λ_i〜3)L_(sd)),B = t_p(L_(ms)=(λ_f 〜3)/(λ_i〜3)L_(sd)),C =(m_n)/ h L_(pm)L_dλ_f,(4)其中h是普朗克常数,m_n是中子质量,λ_i和λ_f是表1中分别给出了入射中子和散射中子,以及其他参数。作者还想感谢Ana Gaspar博士的先前工作,涉及TOFTOF光谱仪及其强度和分辨率函数的估算[14]。 ],将其添加到引用的参考文献列表中。本引证涉及引言,对仪器框架重叠率的讨论(包括图6的标题中的内容)以及对仪器分辨率的讨论。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号