首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >Monte Carlo evaluation of the Filtered Back Projection method for image reconstruction in proton computed tomography
【24h】

Monte Carlo evaluation of the Filtered Back Projection method for image reconstruction in proton computed tomography

机译:质子计算机断层扫描中图像重建的滤波反投影方法的蒙特卡洛评估

获取原文
获取原文并翻译 | 示例
       

摘要

In this paper the use of the Filtered Back Projection (FBP) Algorithm, in order to reconstruct tomographic images using the high energy (200-250 MeV) proton beams, is investigated. The algorithm has been studied in detail with a Monte Carlo approach and image quality has been analysed and compared with the total absorbed dose. A proton Computed Tomography (pCT) apparatus, developed by our group, has been fully simulated to exploit the power of the Geant4 Monte Carlo toolkit. From the simulation of the apparatus, a set of tomographic images of a test phantom has been reconstructed using the FBP at different absorbed dose values. The images have been evaluated in terms of homogeneity, noise, contrast, spatial and density resolution.
机译:本文研究了使用滤波反投影(FBP)算法,以使用高能(200-250 MeV)质子束重建断层图像。已经使用蒙特卡洛方法对算法进行了详细研究,并对图像质量进行了分析,并将其与总吸收剂量进行了比较。我们小组开发的质子计算机断层扫描(pCT)装置已被完全模拟,以利用Geant4蒙特卡洛工具包的功能。通过设备的仿真,已使用FBP在不同的吸收剂量值下重建了测试体模的一组断层图像。已对图像的均一性,噪声,对比度,空间和密度分辨率进行了评估。

著录项

  • 来源
  • 作者单位

    Laboratori Nazionali del Sud - National Instiute for Nuclear Physics INFN (INFN-LNS), Via S.Sofia 64, 95100 Catania, Italy;

    Department of 'Fisiopatologia Clinica', University of Florence, V.le Morgagni 85,1-50134 Florence, Italy;

    Energetic Department, University of Florence, Via S. Marta 3,1-50139 Florence, Italy;

    Laboratorio di Tecnologie Oncologiche HSR, Giglio Contrada, Pietrapollastra-Pisciotto, 90015 Cefalu, Palermo, Italy;

    National Institute for Nuclear Physics INFN, Section of Florence, Via C. Sansone 1, Sesto Fiorentino. 1-50019 Florence, Italy;

    Laboratori Nazionali del Sud - National Instiute for Nuclear Physics INFN (INFN-LNS), Via S.Sofia 64, 95100 Catania, Italy;

    Laboratori Nazionali del Sud - National Instiute for Nuclear Physics INFN (INFN-LNS), Via S.Sofia 64, 95100 Catania, Italy,Nuclear Engineering Department, University of Palermo, Via… Palermo, Italy;

    Physics Department, University of Catania, Via S. Sofia 64,1-95123, Catania, Italy;

    Laboratori Nazionali del Sud - National Instiute for Nuclear Physics INFN (INFN-LNS), Via S.Sofia 64, 95100 Catania, Italy;

    Department of 'Fisiopatologia Clinica', University of Florence, V.le Morgagni 85,1-50134 Florence, Italy;

    National Institute for Nuclear Physics INFN, Section of Catania, Via S.Sofia 64, 95123 Catania, Italy;

    National Institute for Nuclear Physics INFN, Section of Catania, Via S.Sofia 64, 95123 Catania, Italy,Physics Department, University of Catania, Via S. Sofia 64,1-95123, Catania, Italy;

    National Institute for Nuclear Physics INFN, Section of Catania, Via S.Sofia 64, 95123 Catania, Italy;

    Department of 'Fisiopatologia Clinica', University of Florence, V.le Morgagni 85,1-50134 Florence, Italy;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    proton tomography; monte carlo simulation; geant4; reconstruction algorithm;

    机译:质子层析成像蒙特卡洛模拟geant4;重建算法;