首页> 外文期刊>Neurocritical Care >Erroneous intracranial pressure measurements from simultaneous pressure monitoring and ventricular drainage catheters
【24h】

Erroneous intracranial pressure measurements from simultaneous pressure monitoring and ventricular drainage catheters

机译:同时进行压力监测和心室引流导管导致的颅内压力测量错误

获取原文
获取原文并翻译 | 示例
           

摘要

The objective of this report is to highlight the potential for false pressure measurements from systems that combine intracranial pressure (ICP) measurement and ventricular drainage. If the ports of the drain become blocked to the extent that they present a high resistance to cerebrospinal fluid flow, then a significant pressure gradient between the inside and outside of the catheter may be established. Thus, any intracatheter transducer will faithfully record a pressure much lower than true ICP. This holds true for catheter tip transducers when the transducer lies inside the catheter. In the absence of flow, however, pressures will equalize; therefore, accurate measurements may be taken if the drain is temporarily closed. We model this situation and provide simulations of expected measurements in such situations; these compare well to observed clinical readings.
机译:本报告的目的是强调结合颅内压(ICP)测量和心室引流的系统进行假压力测量的可能性。如果排泄口被阻塞到对脑脊液流动具有高抵抗力的程度,则可以在导管的内部和外部之间建立显着的压力梯度。因此,任何导管内换能器将忠实记录远低于真实ICP的压力。当换能器位于导管内部时,这适用于导管尖端换能器。但是,如果没有流量,压力将相等。因此,如果排水管暂时关闭,则可以进行准确的测量。我们对这种情况进行建模,并在这种情况下提供预期测量的模拟;这些与观察到的临床读数比较好。