We present the design, manufacturing, and characterization of a novel miniaturized on-tube-seal configuration for microfkuidic devices. The seal is based on a previously developed world-to-chip spring-based interface by Kortmann et al. (Lab Chip 9:1455-1460, 2009a), which enables rapid and reliable microfluidic connections. In this study, the dead volumes, the contact pressure, the discontinuous fiuidic-profile transmission, and the space requirements were significantly optimized by a new on-tube-seal configuration. Maintaining the advantages of the previously described interface, the new on-tube-seal configuration has a dead volume of only 18 nl, withstands pressures higher than 2,800 kPa with only 3.8 N applied contact force, and allows continuous capillary to chip fluidic profile transmission. The on-tube-seal configuration consists of a miniaturized o-ring (0.5 × 0.3 mm) integrated into a 1/16" tubing that reduces space requirements to a minimal sealing grid of 1.59 mm and is easily adaptable to any planar channel opening of micro fluidic devices. In summary, we present a novel combination of gasket and tubing, which we termed on-tube-seal that allows simple, rapid, and reliable world-to-chip sealing.
机译:我们介绍了microfkuidic设备的新型微型管上密封配置的设计,制造和表征。该密封件基于Kortmann等人先前开发的基于芯片的世界范围的基于弹簧的接口。 (Lab Chip 9:1455-1460,2009a),可实现快速,可靠的微流体连接。在这项研究中,死体积,接触压力,不连续的流体轮廓传输和空间需求都通过新的管上密封配置得到了显着优化。保留了先前描述的接口的优点,新的管上密封构造的死体积仅为18 nl,在仅施加3.8 N的接触力的情况下承受的压力高于2,800 kPa,并允许连续的毛细管到切屑的流体轮廓传输。管上密封结构由集成到1/16英寸管中的微型O形圈(0.5×0.3 mm)组成,可将空间需求减少到1.59 mm的最小密封网格,并易于适应于任何平面通道开口总而言之,我们介绍了垫圈和管道的新颖组合,我们称之为“管上密封”,可实现简单,快速和可靠的世界范围的芯片密封。

著录项

  • 来源
    《Microfluidics and nanofluidics》 |2011年第3期|p.679-684|共6页
  • 作者单位

    Single Cell Laboratory, Leibniz-Institut fur Analytische Wissenschaften - ISAS - e.V., Otto-Hahn-Strasse 6 b, 44227 Dortmund, Germany;

    Single Cell Laboratory, Leibniz-Institut fur Analytische Wissenschaften - ISAS - e.V., Otto-Hahn-Strasse 6 b, 44227 Dortmund, Germany;

    Precision Mechanical Department & Design Department, Leibniz-Institut fur Analytische Wissenschaften - ISAS - e.V., Bunsen-Kirchhoff-Str. 11, 44139 Dortmund. Germany;

    Laboratory of Chemical Biotechnology, TU Dortmund University, 44227 Dortmund, Germany;

    Single Cell Laboratory, Leibniz-Institut fur Analytische Wissenschaften - ISAS - e.V., Otto-Hahn-Strasse 6 b, 44227 Dortmund, Germany,Laboratory of Chemical Biotechnology, TU Dortmund University, 44227 Dortmund, Germany;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    microfluidic; micro seal; world-to-chip interface; single cell analysis; envirostat;

    机译:微流体微密封世界芯片接口;单细胞分析;环保剂;
  • 入库时间 2022-08-17 13:45:35

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号