首页> 外文期刊>Materials science & engineering >Effect of lubricant on the reliability of dental implant abutment screw joint: An in vitro laboratory and three-dimension finite element analysis
【24h】

Effect of lubricant on the reliability of dental implant abutment screw joint: An in vitro laboratory and three-dimension finite element analysis

机译:润滑剂对牙种植体基台螺钉接头可靠性的影响:体外实验室和三维有限元分析

获取原文
获取原文并翻译 | 示例
           

摘要

Biomechanical factors play a key role in the success of dental implants. Fracture and loosening of abutment screws are major issues. This study investigated the effect of lubricants on the stability of dental implant- abutment connection. As lubricants, graphite and vaseline were coated on the abutment screw surface, respec¬ tively, and a blank without lubricant served as the control. The total friction coefficient (μ_(tot)), clamping force, fa¬ tigue behavior and detorque of the joint combined with dynamic cyclic loading were measured under different lubricating conditions. Further, a three-dimensional finite element analysis was used to investigate stress distri¬ bution, in conjunction with experimental images. The results showed that the lubricant reduced μ_(tot), which in turn led to an increase in clamping force. Decrease in loading increased the fatigue life of the screw. However, use of lubricant at high load reduced the fatigue life. Ductile fracture at the first thread of the screw was the chief failure mode, which was due to maximum von Mises stress. Higher stress levels occurred in the lubricant groups. Lubricated screws resulted in lower detorque which made the joint easier to loosen. In conclusion, the lubricant cannot effectively improve the reliability of dental implant-abutment connection. Keeping the inter¬ faces of implant-screw uncontaminated and strengthening the surface of the screw may be recommend for clin¬ ical operation and future design.
机译:生物力学因素在牙科植入物的成功中起关键作用。基台螺丝的断裂和松动是主要问题。这项研究调查了润滑剂对牙种植体-基台连接稳定性的影响。作为润滑剂,分别将石墨和凡士林涂在基台螺钉表面上,将不含润滑剂的坯料作为对照。在不同的润滑条件下,测量了总摩擦系数(μ_(tot)),结合力,关节的疲劳性能和扭矩以及动态循环载荷。此外,结合实验图像,使用三维有限元分析来研究应力分布。结果表明,润滑剂降低了μ_(tot),进而导致夹紧力增加。载荷的减少延长了螺钉的疲劳寿命。但是,在高负荷下使用润滑剂会降低疲劳寿命。螺钉第一螺纹的韧性断裂是主要的破坏模式,这是由于最大的冯·米塞斯应力所致。润滑剂组中的应力水平较高。润滑螺丝可降低扭矩,使接头更容易松开。总之,润滑剂不能有效地提高牙种植体-基台连接的可靠性。对于临床操作和将来的设计,可能建议保持植入物螺钉的接口不受污染并加强螺钉的表面。

著录项

  • 来源
    《Materials science & engineering》 |2017年第6期|297-304|共8页
  • 作者单位

    State Key Laboratory of Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China;

    Department of Biomechanics, Institute of Architecture and Environment, Sichuan University, Chengdu 610065, China;

    State Key Laboratory of Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China;

    State Key Laboratory of Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China;

    State Key Laboratory of Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China;

    State Key Laboratory of Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Dental implant; Lubricant; Fatigue; Finite element analysis; Loosening;

    机译:牙种植体润滑剂疲劳;有限元分析;松动;