Clinical diagnosis of fissure caries with conventional and laser-induced fluorescence techniques

机译:常规和激光诱导荧光技术对裂龋的临床诊断

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We studied the in vivo validity of dentinal fissure caries diagnosis by visual examination, bitewing radiography, and use of a laser-induced fluorescence device (DIAGNOdent). A total of 144 and second molars with macroscopically intact occlusal surfaces in 41 Chinese young adults were examined visually, by bitewing radiography, and by DIAGNOdent. Visual examination after pit and fissure opening was used as the reference standard. The sensitivity and specificity of detecting caries that had extended into the dentin were, respectively, 0.89 and 0.44 by visual detection of opacity or discoloration after air drying, 0.13 and 1.00 by bitewing radiography to detect radiolucency extending into the dentin, and 0.70 and 0.84 by DIAGNOdent testing with a cut-off score of 40. Caries detection by a combination of visual examination and DIAGNOdent had a sensitivity of 0.67 and specificity of 0.94. Receiver operating characteristic analysis showed that this combined approach was superior to the other methods.
机译:我们通过视觉检查,咬合射线照相术以及使用激光诱导荧光装置(DIAGNOdent)研究了齿裂龋的体内有效性。通过咬合射线照相术和DIAGNOdent对41名中国年轻人的宏观上完整的咬合面的144颗和第二颗磨牙进行了视觉检查。凹坑和裂隙开口后的目视检查用作参考标准。通过目视检测风干后的不透明性或变色,检测延伸到牙本质中的龋齿的敏感性和特异性分别为0.89和0.44,通过咬合放射线照相术检测延伸至牙本质中的射线透过性的肉眼检测分别为0.13和1.00,以及通过放射线照相术检测延伸至牙本质中的放射线透明度的敏感性和特异性为0.70和0.84。 DIAGNOdent检测的满分为40。通过视觉检查和DIAGNOdent结合进行龋齿检测的敏感性为0.67,特异性为0.94。接收机工作特性分析表明,这种组合方法优于其他方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号