首页> 外文期刊>Journal of Radioanalytical and Nuclear Chemistry >Calibration of the HB-line active well neutron coincidence counter for measure of LANL 3013 highly enriched uranium product splits
【24h】

Calibration of the HB-line active well neutron coincidence counter for measure of LANL 3013 highly enriched uranium product splits

机译:HB线活动井中子符合计数器的校准,用于测量LANL 3013高浓缩铀产品分裂

获取原文
获取原文并翻译 | 示例
           

摘要

In this paper, the setup, calibration, and testing of the F-Area Analytical Labs active well neutron coincidence counter (HV-221000-NDA-X-1-DK-AWCC-1) in SRNL are described for use in the Savannah River Site (SRS) transuranium metal production facility to enable assay of mixed uranium/plutonium metal product. The instrument was required within a three-month window for availability upon receipt of LANL uranium oxide samples into the SRS facility. Calibration of the instrument in the SRNL nuclear nondestructive assay facility in the range 10–400 g HEU is described. We also report qualification and installation of the instrument for assay of the initial suite of product samples.
机译:在本文中,描述了在萨凡纳河中使用的SRNL中F区域分析实验室活动井中子重合计数器(HV-221000-NDA-X-1-DK-AWCC-1)的设置,校准和测试站点(SRS)超铀金属生产设施,可用于分析混合铀/ mixed金属产品。收到LANL氧化铀样品进入SRS设施后,需要在三个月的时间内提供该仪器。描述了在SRNL核无损分析设备中仪器的校准范围为10–400 g HEU。我们还将报告用于鉴定初始产品样品套件的仪器的鉴定和安装情况。