摘要

A Prompt Gamma Ray Neutron Activation Analysis (PGNAA) system, incorporating an isotopic neutron source has been simulated using the MCNPX Monte Carlo code. In order to improve the signal to noise ratio different collimators and a filter were placed between the neutron source and the object. The effect of the positioning of the neutron beam and the detector relative to the object has been studied. In this work the optimisation procedure is demonstrated for boron. Monte Carlo calculations were carried out to compare the performance of the proposed PGNAA system using four different neutron sources (241Am/Be, 252Cf, 241Am/B, and DT neutron generator). Among the different systems the 252Cf neutron based PGNAA system has the best performance.
机译:使用MCNPX蒙特卡洛代码对包含同位素中子源的瞬发伽马射线中子活化分析(PGNAA)系统进行了模拟。为了提高信噪比,在中子源和物体之间放置了不同的准直器和滤波器。已经研究了中子束和探测器相对于物体的定位效果。在这项工作中,硼的优化程序得到了证明。使用四种不同的中子源(241 Am / Be,252 Cf,241 Am / B和DT中子发生器)进行了蒙特卡罗计算,以比较拟议的PGNAA系统的性能。 )。在不同的系统中,基于252 Cf中子的PGNAA系统具有最佳性能。