首页> 外文期刊>Journal of Mathematical Imaging and Vision >Sampling Bounds for 2-D Vector Field Tomography
【24h】

Sampling Bounds for 2-D Vector Field Tomography

机译:二维矢量场断层扫描的采样范围

获取原文
获取原文并翻译 | 示例
           

摘要

The tomographic mapping of a 2-D vector field from line-integral data in the discrete domain requires the uniform sampling of the continuous Radon domain parameter space. In this paper we use sampling theory and derive limits for the sampling steps of the Radon parameters, so that no information is lost. It is shown that if Δx is the sampling interval of the reconstruction region and x max is the maximum value of domain parameter x, the steps one should use to sample Radon parameters ρ and θ should be: and . Experiments show that when the proposed sampling bounds are violated, the reconstruction accuracy of the vector field deteriorates. We further demonstrate that the employment of a scanning geometry that satisfies the proposed sampling requirements also increases the resilience to noise.
机译:离散域中来自线积分数据的二维矢量场的层析成像需要连续Radon域参数空间的均匀采样。在本文中,我们使用采样理论并推导Radon参数采样步骤的限制,从而不会丢失任何信息。结果表明,如果Δx是重构区域的采样间隔,而x max 是域参数x的最大值,则采样Radon参数ρ和θ所应采用的步骤为:和。实验表明,当违反建议的采样范围时,矢量场的重构精度会下降。我们进一步证明,使用满足建议的采样要求的扫描几何形状还可以提高抗噪能力。