Laser cutting and wire forming are two of the most commonly used processes in the manufacture of Nitinol medical devices. This study explores how varying the amount of material removed during the final surface treatment steps affects the corrosion resistance of Z-type stents that have either been laser-cut from tube or shape set from wire. All parts were subjected to a typical heat treatment process necessary to achieve an Austenite finish (Af) temperature of 25 ± 5 °C, and were subsequently post-processed with an electrochemical passivation process. The total weight loss during post-processing was recorded and the process adjusted to create groups with less than 5%, less than 10%, and less than 25% amounts of weight loss. The parts were then crimped to 6 mm and allowed to expand back to their original diameter. The corrosion test results showed that on average both groups of Z-stents experienced an increase in the corrosion breakdown potential and a decrease in the standard deviation with increasing amounts of material removal. In addition, less material removal is required from the wire-form Z-stents as compared to the laser-cut Z-stents to achieve high corrosion resistance. Finally, 7 day nickel ion release tests performed on the wire-formed Z-stents showed a dramatic decrease from 0.0132 mg of nickel leached per day for the low weight loss group to approximately 0.001 mg/day for the medium and high weight loss groups.
机译:激光切割和金属丝成型是镍钛诺医疗设备制造中最常用的两种工艺。这项研究探索了在最后的表面处理步骤中改变材料去除量如何影响Z型支架的耐腐蚀性,该Z型支架是从管子上激光切割或从金属丝上定型的。所有零件均经过典型的热处理工艺,以达到25±5°C的奥氏体表面温度(A ),然后进行电化学钝化处理。记录后处理过程中的总重量损失,并对过程进行调整,以创建重量减少量少于5%,少于10%和少于25%的组。然后将零件压接至6 mm,然后使其膨胀回原来的直径。腐蚀测试结果表明,随着材料去除量的增加,两组Z支架平均都经历了腐蚀击穿电位的增加和标准偏差的减小。此外,与激光切割的Z型支架相比,从丝状Z型支架中去除材料的要求更少,以实现高耐腐蚀性。最后,在线状Z型支架上进行的7天镍离子释放测试表明,从低重量减轻组每天浸出的镍0.0132 mg /天到中等和高重量减轻组每天大约0.001 mg /天的显着降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号