Riata lead failure with normal electrical lead parameters and normal fluoroscopic appearance

机译:Riata导线失效,具有正常的电导线参数和正常的荧光检查外观

获取原文
获取原文并翻译 | 示例
           

摘要

Recently, insulation breach of the Riata lead raised a big concern. Management of these externalized leads has been addressed by professional organizations. However, what to do in patients with Riata leads without manifested failure is an ongoing clinical dilemma. Here, we present two clinical scenarios where the implantable cardioverter defibrillator system failed to deliver shock therapy in spite of having a new ICD generator and “appropriately functioning leads” as revealed by lead interrogation.
机译:最近,Riata铅的绝缘破坏引起了极大关注。这些外部线索的管理已由专业组织解决。然而,在没有明显失败的情况下对Riata导联患者进行治疗是一个持续的临床难题。在这里,我们介绍了两种临床情况,其中植入式心脏复律除颤器系统尽管配备了新的ICD发生器和“适当运行的导线”,但仍无法进行电击疗法,如导线询问所揭示的那样。