首页> 外文期刊>Journal of Huazhong University of Science and Technology >Development and Clinical Application of An Quantitative Head-Band Formed Ocular Compressor
【24h】

Development and Clinical Application of An Quantitative Head-Band Formed Ocular Compressor

机译:定量头带式眼压治疗仪的研制与临床应用

获取原文
获取原文并翻译 | 示例
       

摘要

In order to investigate the efficiency of a new quantitative head-band formed ocular compressor to reduce intraocular pressure (IOP), ocular compression by this new reducer with 40 mmHg for 10 mm was performed on 87 cataractous eyes of 78 cases. The changes of IOP (87 eyes) and anterior chamber depth (ACD) were observed. There was a significant decrease of IOP and increase of ACD within 30 min after decompression (P<0. 001). The mean decrease of IOP was 5. 62+-2. 41 mmHg and the mean increase of ACD was 0. 18+-0. 09 mm within 5 min after decompression. The IOP 5 min after decompression had no significant difference with that 10 min after decompression (P >0. 05). IOP below 10 mmHg could last for about 15 min. This apparatus had been successfully applied to 80 eyes for extracapsular cataract extraction. It was suggested that this device had the advantages of safety, accurate quantification, reliable effect, casually adjusting pressure according to various demands and time-saving.
机译:为了研究新型定量头带式眼用压缩器降低眼内压(IOP)的效率,使用这种新型40 mmHg的减压器对10毫米眼压进行了眼压,治疗了78例患者。观察到眼压(87眼)和前房深度(ACD)的变化。减压后30分钟内,眼压显着降低,ACD升高(P <0.001)。 IOP的平均下降为5。62 + -2。 41 mmHg和ACD的平均增加为0. 18 + -0。减压后5分钟内达到09毫米。减压后5分钟的眼压与减压后10分钟的眼压没有显着差异(P> 0。05)。低于10 mmHg的IOP可能持续约15分钟。该仪器已成功应用于80眼囊外白内障摘除术。建议该装置具有安全,定量准确,效果可靠,根据各种需求随意调节压力,节省时间的优点。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号