首页> 外文期刊>Journal of Heat Transfer >Subtle Issues in the Measurement of the Thermal Conductivity of Vacuum Insulation Panels
【24h】

Subtle Issues in the Measurement of the Thermal Conductivity of Vacuum Insulation Panels

机译:真空隔热板导热系数测量中的细微问题

获取原文
获取原文并翻译 | 示例
           

摘要

Vacuum insulation panels have values of thermal conductivity that are extremely low (~4 mW/m · K) compared to the thermal conductivity of most common insulations. Typical ASTM test methods are not designed for testing these very low thermal conductivity materials. An apparatus has been built and tested that uses a thin foil heater and vacuum chamber to test vacuum insulation panels. Several different measurement configurations are studied to determine the effects of the parasitic heat losses. The differences between the ASTM standards and this technique are described and the rationale explained. A new ASTM technique for vacuum panels appears to be needed.
机译:真空隔热板的导热系数与大多数普通隔热板的导热系数相比非常低(〜4 mW / m·K)。典型的ASTM测试方法并非设计用于测试这些导热系数非常低的材料。已经建立并测试了一种使用薄箔加热器和真空室来测试真空隔热板的设备。研究了几种不同的测量配置,以确定寄生热量损失的影响。描述并解释了ASTM标准与该技术之间的差异。似乎需要一种用于真空面板的新ASTM技术。