首页> 外文期刊>Journal of Endourology >Rationale and Feasibility Study of a Mechanical Model for the Testing of Material Fatigue in Metal Ureteral Stents
【24h】

Rationale and Feasibility Study of a Mechanical Model for the Testing of Material Fatigue in Metal Ureteral Stents

机译:金属输尿管支架材料疲劳测试力学模型的理论基础和可行性研究

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Stents are used abundantly to maintain ureteral patency. The majority are plastic tubes that adjust easily to upper urinary-tract motion. Recently, a coiled-wire lumenless stent was introduced (ZebraStent®, Neo Medical, Munich, Germany) to facilitate expulsion of stone fragments after lithotripsy. Its metal core is composed of Nitinol, with the soft J ends being of titanium. The thin shape considerably increases the extraluminal space. The ZebraStent stretches the ureter and also provides a surface for the fragments to glide along. In our 18-month experience with the ZebraStent, two of them fractured along the shaft. We sought to learn whether this complication resulted from a defect in stent design or from material fatigue secondary to constant movement.nnMaterials and Methods: Our model is powered by an electric motor that produces a constant displacement similar to stent movements in vivo. The whole ZebraStent is embedded in a 37°C waterbath to simulate physiological conditions within the ureter. We used an average displacement of 16 mm. The average frequency of ventilatory-cycle simulation was 20 times that in vivo, allowing us to collect data in a shorter time.nnResults: All 10 stents broke within the proximal Nitinol shaft at the equivalent of 4 to 6 months (125–179 days).nnConclusions: Our preliminary results show that all stents break after the equivalent of 4 or more months. The fact that this occurs in the homogenous proximal Nitinol shaft rather than at the welding point between the shaft and the titanium curl implies that breakage is secondary to material fatigue and not design error. Extensive testing is under way to confirm material fatigue as the cause of breakage. We hope to determine a safe dwelling time for these stents, which at the moment should not exceed 3 months.
机译:背景:大量使用支架来维持输尿管通畅。大多数是塑料管,可以很容易地适应上尿路的运动。最近,引入了一种无线圈螺旋腔支架(ZebraStent®,Neo Medical,慕尼黑,德国),以促进碎石后结石碎片的排出。它的金属芯由镍钛诺组成,柔软的J端由钛制成。薄的形状大大增加了腔外空间。 ZebraStent可以拉伸输尿管,还可以为碎片提供顺滑的表面。在我们使用ZebraStent的18个月经验中,其中两个沿轴断裂。我们试图了解这种并发症是由于支架设计缺陷还是由于持续运动引起的材料疲劳所致。nn材料和方法:我们的模型由电动马达驱动,该马达产生与体内支架运动相似的恒定位移。整个ZebraStent均嵌入37°C水浴中,以模拟输尿管内的生理状况。我们使用的平均位移为16毫米。通气循环模拟的平均频率是体内的20倍,因此我们可以在更短的时间内收集数据。nn结果:所有10个支架在镍钛合金近端干系内断裂的时间为4至6个月(125-179天) .nn结论:我们的初步结果表明,所有支架均在相当于4个月或更长时间的情况下破裂。这种情况发生在均匀的镍钛合金近端轴中,而不是在轴和钛卷曲之间的焊接点处,这一事实表明,断裂是材料疲劳的继发因素,而不是设计错误。正在进行广泛的测试以确认材料疲劳是断裂的原因。我们希望为这些支架确定一个安全的居住时间,目前不应超过3个月。

著录项

  • 来源
    《Journal of Endourology》 |2008年第2期|p.389-392|共4页
  • 作者单位

    Nikos Bafaloukas, M.D., Ph.D.Department of Urology, Barts and the London NHS Trust, London, United Kingdom.Malcom Birch, B.Eng., M.Sc., Ph.D.Department of Clinical Physics, Barts and the London NHS Trust, London, United Kingdom.Noor Buchholz, M.D.Department of Urology, Barts and the London NHS Trust, London, United Kingdom.;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Stents,ureteral,urinary-tract motion;

    机译:支架;输尿管;尿路运动;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号