An argon-nitrogen-hydrogen mixed-gas plasma was optimized towards the minimization of matrix effects in inductively coupled plasma optical emission spectrometry. Analyte sensitivity was enhanced by a factor of 1 to 4 (n = 26 elements) upon adding 3.7% v/v N_2 to the plasma gas and 1.1% v/v H_2 as a sheath around the Ar nebulizer gas flow in the central channel. However, detection limits were on average similar when using the mixed-gas plasma versus the Ar plasma, although at least 2-fold improvements resulted for Al and Ca in presence of 5% m/v Na. The 8.9 ± 1.7% (n = 26 elements) analyte signal suppression with an Ar plasma in a 0.23% m/v (or 0.1 M) Na matrix was reduced to 5.6 ± 2.3% with the mixed-gas plasma. The mixed-gas plasma was significantly more robust, with a Mg II/Mg I ratio of 13.3 ± 0.1 (n = 5 replicates), compared to 9.87 ± 0.16 for an Ar plasma optimized for robustness. The plasma excitation temperature also increased by 717 K for the mixed-gas plasma, confirming the improved excitation conditions. The superior analytical performance of the robust mixed-gas plasma for complex sample matrices was demonstrated through the direct quantitative multi-elemental analysis of a variety of certified reference materials (i.e., lake sediment, till, stream sediment, and natural ore digests), without internal standardization or matrix-matched calibration.
机译:优化了氩-氮-氢混合气体等离子体,以使感应耦合等离子体光发射光谱法中的基质效应最小化。在向中央通道的Ar雾化器气流周围添加3.7%v / v N_2和1.1%v / v H_2作为鞘时,分析物灵敏度提高了1-4倍(n = 26个元素)。但是,使用混合气体等离子体与Ar等离子体时,检测限平均相似,尽管在存在5%m / v Na的情况下,Al和Ca至少提高了2倍。在混合气体等离子体中,在0.23%m / v(或0.1 M)的Na基体中使用Ar等离子体,分析物信号抑制8.9±1.7%(n = 26个元素),降至5.6±2.3%。混合气体等离子体具有更高的鲁棒性,Mg II / Mg I比为13.3±0.1(n = 5次重复),而针对鲁棒性进行了优化的Ar等离子体则为9.87±0.16。对于混合气体等离子体,等离子体激发温度也增加了717 K,从而确认了改善的激发条件。通过对多种认证参考物质(例如,湖泊沉积物,耕种,河流沉积物和天然矿石消化物)进行直接定量多元素分析,证明了鲁棒的混合气体等离子体对复杂样品基质的出色分析性能。内部标准化或矩阵匹配的校准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号