Investigation of a Systematic Offset in the Measurement of Organic Carbon with a Semicontinuous Analyzer

机译:半连续分析仪测量有机碳中的系统偏移的研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Organic carbon (OC) was measured semicontinuously in laboratory experiments of steady-state secondary organic aerosol formed by hydrocarbon + nitrogen oxide irradiations. Examination of the mass of carbon measured on the filter for various sample volumes reveals a systematic offset that is not observed when performing an instrumental blank. These findings suggest that simple subtraction of instrumental blanks determined as the standard analysis without sample collection (i.e., by cycling the pump and valves yet filtering zero liters of air followed by routine chemical analysis) from measured concentrations may be inadequate. This may be especially true for samples collected through the filtration of small air volumes wherein the influence of the systematic offset is greatest. All of the experiments show that filtering a larger volume of air minimizes the influence of contributions from the systematic offset. Application of these results to measurements of ambient concentrations' of carbonaceous aerosol suggests a need for collection of sufficient carbon mass to minimize the relative influence of the offset signal.
机译:在由碳氢化合物+氮氧化物辐照形成的稳态二次有机气溶胶的实验室实验中,半连续测量了有机碳(OC)。对各种样品量在过滤器上测得的碳质量的检查显示出系统的偏移,该偏移在执行仪器空白时没有观察到。这些发现表明,从测量浓度中简单减去作为标准分析而确定的仪器空白(例如,通过循环泵和阀门,但过滤零升空气然后进行常规化学分析)可能是不够的。对于通过小风量过滤收集的样品尤其如此,其中系统偏移的影响最大。所有的实验都表明,过滤更多的空气可以最大程度地减少系统补偿的影响。将这些结果应用于碳质气溶胶环境浓度的测量表明,需要收集足够的碳质量以最小化偏移信号的相对影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号