首页> 外文期刊>Isotopes in environmental and health studies >The first IAEA inter-laboratory comparison exercise in Latin America and the Caribbean for stable isotope analyses of water samples
【24h】

The first IAEA inter-laboratory comparison exercise in Latin America and the Caribbean for stable isotope analyses of water samples

机译:拉丁美洲和加勒比地区的第一个国际原子能机构实验室比较运动,用于稳定同位素的水样

获取原文
获取原文并翻译 | 示例
           

摘要

The use of stable isotopes (delta H-2 and delta O-18) is widespread in water resources studies. In the Latin America and the Caribbean (LAC) region, the application of isotope techniques has increased in the past decade, but there remains room to gain self-reliance in environmental isotope studies, necessitating easy and fast access to good-quality isotope data. To that end, in 2018 the IAEA carried out the first regional interlaboratory comparison exercise, testing the analytical performance of 25 laboratories using isotope-ratio mass spectrometry and laser absorption spectroscopy. The three test samples covered a commonly observed range of 0 to -16 parts per thousand delta O-18 and 0 to -115 parts per thousand delta H-2. z- and zeta-scores were used to benchmark laboratories' performance against a strict criterion. We found that 81% of the laboratories had satisfactory performance (vertical bar z vertical bar = 2) for delta H-2 but only 54% achieved similar scores for delta O-18. Only a minor fraction of results (12% for delta H-2 and 15% for delta O-18) were unsatisfactory. The larger number of questionable results for delta O-18 confirmed the challenges in laser absorption spectroscopy for this isotope. Besides instrumental performance, the sample throughput, laboratory reference materials, and data post-processing were contributing factors to inaccurate or imprecise performance.
机译:使用稳定同位素(Delta H-2和Delta O-18)是水资源研究的广泛。在拉丁美洲和加勒比地区(LAC)地区,在过去十年中,同位素技术的应用增加,但仍有余地在环境同位素研究中获得自依赖性,需要轻松和快速地获得高质量的同位素数据。为此,2018年,原子能机构进行了第一个区域间双向化比较运动,使用同位素比质谱和激光吸收光谱检测25实验室的分析性能。三个测试样品覆盖了每千次δO-18和0至-115份0至-115份每千次δH-2的常见范围。 Z-和Zeta分数用于根据严格的标准来基准实验室的表现。我们发现,81%的实验室具有令人满意的性能(垂直条Z垂直条<= 2),但仅为Delta O-18的相似评分达到了54%。只有少量的结果(δH-2的12%和δO-18的15%)都是不令人满意的。 Delta O-18的更多可疑结果结果证实了这种同位素激光吸收光谱的挑战。除了乐器性能,样品吞吐量,实验室参考资料和数据后处理是否有助于不准确或不精确表现的因素。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号