An interlaboratory comparison using relative-humidity (RH) and temperature probes at three national measurement institutes and two accredited laboratories has been carried out. The work had three purposes: firstly, to establish the instruments’ level of reproducibility and suitability for use as transfer standards within their specified range of operation; secondly, to show the agreement of a method of RH generation utilizing certified non-saturated salt RH standards when compared with a method of RH calibration using a chilled-mirror reference and platinum-resistance thermometers; and finally, from the results obtained it is possible to establish the equivalence between the participating laboratories, to the level of uncertainty achievable with the transfer standards used. A total of six RH probes were tested in two groups. The instruments of the first group were calibrated in the range from 10 %rh to 90 %rh at a temperature of 23 °C. The second group of instruments was measured in the same RH range, but at the temperatures of 5 °C, 23 °C, and 50 °C. The objective of the tests on the second group of instruments was to determine the effect of a wider operating temperature range on performance. This article presents and discusses the results of the comparison in the context of an international collaboration that provides confidence in the measurements performed by the participants within their respective accredited scopes and the ILAC or the CIPM mutual recognition arrangements.
机译:在三个国家测量机构和两个认证实验室进行了使用相对湿度(RH)和温度探针的实验室间比较。这项工作具有三个目的:首先,确定仪器在指定操作范围内的再现性和适用性,以用作转移标准。其次,与使用冷镜参考和铂电阻温度计进行RH校准的方法相比,要表明使用认证的非饱和盐RH标准进行RH的方法的一致性;最后,根据获得的结果,有可能确定参与实验室之间的等效性,以达到所使用转移标准所能达到的不确定性水平。两组共测试了六个RH探针。第一组的仪器在23°C的温度下在10%rh至90%rh的范围内校准。第二组仪器是在相同的相对湿度范围内,但在5°C,23°C和50°C的温度下测量的。在第二组仪器上进行测试的目的是确定更宽的工作温度范围对性能的影响。本文介绍并讨论了在国际合作中进行比较的结果,该合作使参与者对在各自认可的范围内以及ILAC或CIPM互认安排下进行的测量充满信心。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号