Quantitative ion mobility spectroscopy based on molecular collision rate theory

机译:基于分子碰撞速率理论的定量离子迁移谱

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Atmospheric pressure chemical ionization and ion mobility spectrometry (IMS) have traditionally been viewed as a qualitative analytical technique for identifying specific chemicals in the atmosphere. This work employs a nonlinear model based on molecular collision rate theory for quantitative modeling of chemical analyte concentrations. The collision rate between any two molecules depends on the relative populations of each chemical species in the volume of air analyzed where most collisions between ions, or neutral molecules and ions, result in no charge transfer. The rate constants for formation of product ions and consumption of source ions are estimated using empirical data over a wide concentration range for several analytes and reagent gases. The rate constants are unique to the analyte and the reagent gas as well as the sensitivity of the particular IMS instrument and provide a quantitative model to relate the mobility peak amplitudes to the analyte concentration. The rate constants can also be normalized by the reaction ion consumption rate constant to remove the IMS instrument sensitivity and provide a qualitative metric for analyte identification independent of a particular IMS instrument. A quantitative example is given for an acetic acid plume measured by a hand-held IMS detector outdoors has the plume passes. The quantitative rate constants provide a reasonable basis for estimating analyte concentration from the ion mobility spectra over a wide range of analyte concentrations.
机译:传统上,大气压化学电离和离子迁移谱(IMS)被视为用于鉴定大气中特定化学物质的定性分析技术。这项工作采用基于分子碰撞速率理论的非线性模型对化学分析物浓度进行定量建模。任何两个分子之间的碰撞速率取决于所分析的空气中每种化学物质的相对种群,其中离子(或中性分子与离子)之间的大多数碰撞不会导致电荷转移。使用多种分析物和反应气在宽浓度范围内的经验数据估算产物离子形成和源离子消耗的速率常数。速率常数对于分析物和反应气以及特定IMS仪器的灵敏度而言是唯一的,并且提供了定量模型以将迁移率峰幅度与分析物浓度相关联。速率常数也可以通过反应离子消耗速率常数进行归一化,以消除IMS仪器的灵敏度,并提供独立于特定IMS仪器的用于分析物鉴定的定性指标。举了一个定量的例子,说明了在室外羽流通过的情况下,由手持式IMS检测器测量的乙酸羽流。定量速率常数为在较宽的分析物浓度范围内根据离子迁移谱估算分析物浓度提供了合理的基础。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号