Can thermocouple measure surface temperature of light emitting diode module accurately?

机译:热电偶可以准确测量发光二极管模块的表面温度吗?

获取原文
获取原文并翻译 | 示例
       

摘要

Thermocouple is often used to measure surface temperature of light emitting diode (LED). However, it was found that the method was not accurate in this work. An experiment was conducted to prove this in the first part. In the experiment, the air temperature above a lit LED chip was measured by a thermocouple, and it was found the measurement temperatures were higher than the LED junction temperatures. When the drive current was 1000 mA, the deviation reached as much as 148.61%, which proved that the temperature measured by the thermocouple was not accurate. The reason for the inaccuracy was supposed to be the absorption of the optical energy by the thermocouple. Theoretical derivation was conducted and it was found that the measurement temperature was related with the local luminance around the thermocouple and the optical absorption coefficient of the thermocouple. Comparison experiments were conducted and qualitatively validated the supposition.
机译:热电偶通常用于测量发光二极管(LED)的表面温度。但是,发现该方法在这项工作中不准确。在第一部分中进行了实验以证明这一点。在实验中,通过热电偶测量点亮的LED芯片上方的空气温度,发现测量温度高于LED结温。当驱动电流为1000 mA时,偏差高达148.61%,这证明由热电偶测得的温度是不准确的。误差的原因可能是由于热电偶吸收了光能。进行了理论推导,发现测量温度与热电偶周围的局部亮度和热电偶的光吸收系数有关。进行比较实验并定性验证了该假设。