Inductively coupled plasma mass spectrometry (ICP-MS) was used for investigating iodine in deep geological repository barriers. Major problems that can arise in the method application include different behaviours of individual iodine species, memory effects and strong nonspectral interferences owing to high salt content in samples. When proper conditions were applied (i.e. the addition of tetramethylammonium hydroxide, use of Ge as an internal standard, and spiking of iodate calibration solutions with sample matrix), ICP-MS provided accurate results, irrespective of the iodine speciation. The quantification limit of iodine determination in highly saline samples was 1.1 μg/L I. The calibration was linear at least within the range of 1-100 μg/L I. The analysis accuracy was verified by an analysis of a model sample of synthetic bentonite water and a 0.1 mol/L NaCl matrix that contained various iodine species (iodide, elemental iodine and iodate) and their mixtures, and recovery ranged from 97 to 101%. The analysis by ICP-MS coupled with liquid chromatography (LC) could also distinguish between the main iodine species of iodide and iodate. The detection limits were less than 2.0 μg/L I for both species. Both ICP-MS determination of the total iodine and the LC-ICP-MS speciation analysis were applied to a preliminary investigation of iodine diffusion through Czech bentonite from the Rokle deposit. No iodide speciation changes were observed during the diffusion experiment which was performed under atmospheric conditions.
机译:电感耦合等离子体质谱法(ICP-MS)用于研究深层地质储层壁垒中的碘。该方法应用中可能出现的主要问题包括各个碘物种的行为不同,记忆效应以及由于样品中盐含量高而引起的强烈的非光谱干扰。当采用适当的条件时(即添加四甲基氢氧化铵,使用Ge作为内标,以及将碘酸盐校准溶液与样品基质加标),无论碘形态如何,ICP-MS都能提供准确的结果。高盐度样品中碘的定量限为1.1μg/ LI。校准至少在1-100μg/ L I范围内是线性的。分析准确性通过合成膨润土模型样品的分析得到验证。水和0.1 mol / L NaCl基质,其中包含各种碘物质(碘化物,元素碘和碘酸盐)及其混合物,回收率在97%至101%之间。 ICP-MS结合液相色谱(LC)进行的分析还可以区分主要的碘化物类型的碘化物和碘酸盐。两种物质的检出限均小于2.0μg/ LI。 ICP-MS测定总碘和LC-ICP-MS形态分析都用于碘从Rokle矿床通过捷克膨润土扩散的初步研究。在大气条件下进行的扩散实验期间,未观察到碘化物的形态变化。

著录项

  • 来源
  • 作者单位

    Department of Analytical Chemistry, Institute of Chemical Technology, Prague, Czech Republic;

    Department of Analytical Chemistry, Institute of Chemical Technology, Prague, Czech Republic;

    Department of Analytical Chemistry, Institute of Chemical Technology, Prague, Czech Republic;

    Department of Nuclear Chemistry, Czech Technical University in Prague, Prague, Czech Republic,UJV Rez, Corp., Husinec - Rez, Czech Republic;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    bentonite; HPLC; ICP-MS; iodine; speciation;

    机译:膨润土HPLC;ICP-MS;碘;物种;
  • 入库时间 2022-08-17 13:30:43

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号