机译:频率范围为10至100 MHz的衰减测量系统

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

An attenuation measurement system, which employs a traceable inductive voltage divider (IVD) at 1 kHz as a reference standard and dual-channel intermediate frequency (IF) substitution method, is developed as an attenuation standard in the frequency range of 10 to 100 MHz. High isolation between the channels is achieved by introducing an assembly of light emitting diode, optical fiber, and photodetector into the system. Sources of uncertainties of the system, such as leakage, are discussed and the expanded uncertainties are expected to be 0.001 dB for an attenuation range up to 20 dB and 0.04 dB for 80 dB.
机译:开发了一种衰减测量系统,它以1 kHz的可追溯电感分压器(IVD)作为参考标准,并采用双通道中频(IF)替代方法,作为10至100 MHz频率范围内的衰减标准。通过将发光二极管,光纤和光电探测器的组件引入系统,可以实现通道之间的高度隔离。讨论了系统不确定性的来源,例如泄漏,对于最大20 dB的衰减范围,扩展的不确定性预计为0.001 dB,对于80 dB的扩展不确定度为0.04 dB。