Calibration of RF field strength meters

机译:射频场强仪的校准

获取原文
获取原文并翻译 | 示例
           

摘要

An intercomparison of field strength meters was performed during 1991-1993 over the frequency range 100 kHz-1000 MHz and field strength range 1-10 V/m using different calibration techniques. Four field sensors, electric as well as magnetic, have been used. Seven institutes participated in this intercomparison using at least three different calibration methods. A calibration uncertainty of less than 1 dB can be obtained if certain procedures are followed. The results are applicable in the fields of EMC testing and health care (nonionizing radiation).
机译:使用不同的校准技术,在1991-1993年期间在100 kHz-1000 MHz频率范围和1-10 V / m的场强范围内进行了场强计的比较。已经使用了四个场传感器,包括电的和磁的。七个研究所使用至少三种不同的校准方法参加了这种比较。如果遵循某些步骤,则可获得的校准不确定度小于1 dB。结果适用于EMC测试和医疗保健(非电离辐射)领域。