首页> 外文期刊>IEEE Transactions on Instrumentation and Measurement >Attenuation-Measurement System in the Frequency Range of 18–40 GHz
【24h】

Attenuation-Measurement System in the Frequency Range of 18–40 GHz

机译:18–40 GHz频率范围内的衰减测量系统

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The attenuation-measurement system of the National Metrology Institute of Japan - the dual-channel intermediate-frequency (IF)-substitution method using an inductive-voltage divider (IVD) as a reference standard - is improved to cover the frequency measurement range of 18-40 GHz. Higher accuracy is obtained by raising the IF to 10 kHz and employing two seven-decade IVDs in cascade arrangement. The use of the frequency doublers at each local port of the mixers allows the low-pass filters to be installed to satisfy the essential isolation between the channels. A 10-kHz voltage-ratio standard has been constructed and used as a reliable technique for uncertainty evaluation and long-term accuracy control of the 10-kHz null-balance receiver. The expanded uncertainties are expected to be 0.002 dB for attenuation up to 40 dB, 0.005 dB for 60 dB, and 0.03 dB for 80 dB
机译:日本国立计量学会的衰减测量系统-以感应分压器(IVD)为参考标准的双通道中频(IF)替代方法-改进以覆盖18的频率测量范围-40 GHz。通过将IF提升至10 kHz并采用两个串联的七个七次IVD,可以获得更高的精度。在混频器的每个本地端口使用倍频器,可以安装低通滤波器,以满足通道之间的基本隔离要求。已经构建了10kHz的电压比标准,并将其用作可靠的技术,用于不确定性评估和10kHz零平衡接收器的长期精度控制。对于最大40 dB的衰减,扩展的不确定度预计为0.002 dB,对于60 dB,0.005 dB,对于80 dB,0.03 dB。