Interoperator Reproducibility of Carotid Elastography for Identification of Vulnerable Atherosclerotic Plaques

机译:颈动脉弹性成像术的互操作性可重复性,用于识别易损的动脉粥样硬化斑块。

获取原文
获取原文并翻译 | 示例
           

摘要

Ultrasound-based carotid elastography has been developed to evaluate the vulnerability of carotid atherosclerotic plaques. The aim of this study was to investigate the in vivo interoperator reproducibility of carotid elastography for the identification of vulnerable plaques, with high-resolution magnetic resonance imaging (MRI) as reference. Ultrasound radiofrequency data of 45 carotid arteries (including 53 plaques) from 32 volunteers were acquired separately by two experienced operators in the longitudinal view and then were used to estimate the interframe axial strain rate (ASR) with a two-step optical flow method. The maximum 99th percentile of absolute ASR of all plaques in a carotid artery was used as the elastographic index. MRI scanning was also performed on each volunteer to identify the vulnerable plaque. The results showed no systematic bias in the Bland-Altman plot and an intraclass correlation coefficient of 0.66 between the two operators. In addition, no statistical significance was found between the receiver operating characteristic (ROC) curves from the two operators (p > 0.05), and their areas under the ROC curves were 0.83 and 0.77, respectively. Using the mean measurements of the two operators as the classification criterion, a sensitivity of 71.4%, a specificity of 87.1%, and an accuracy of 82.2% were obtained with a cutoff value of 1.37 s(-1). This study validates the interoperator reproducibility of ultrasound-based carotid elastography for identifying vulnerable carotid plaques.
机译:已经开发了基于超声的颈动脉弹性成像技术,以评估颈动脉粥样硬化斑块的脆弱性。这项研究的目的是调查高分辨率的磁共振成像(MRI)在颈动脉弹性成像的体内互操作性重现性鉴定易损斑块。两名经验丰富的操作员在纵向上分别从32名志愿者中采集了45个颈动脉(包括53个斑块)的超声射频数据,然后使用两步光流法将其用于估计框架间轴向应变率(ASR)。颈动脉中所有斑块的绝对ASR的最大99%用作弹性成像指标。还对每位志愿者进行了MRI扫描以识别易损斑块。结果表明在Bland-Altman图中没有系统偏差,并且两个算子之间的类内相关系数为0.66。此外,在两个操作员的接收器工作特性(ROC)曲线之间未发现统计显着性(p> 0.05),它们在ROC曲线下的面积分别为0.83和0.77。使用两个算子的平均测量值作为分类标准,灵敏度值为71.4%,特异性为87.1%,准确度为82.2%,临界值为1.37 s(-1)。这项研究验证了基于超声的颈动脉弹性成像术的互操作性可重复性,可用于识别脆弱的颈动脉斑块。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号