首页> 外文期刊>Robotics, IEEE Transactions on >Design of an Optically Controlled MR-Compatible Active Needle
【24h】

Design of an Optically Controlled MR-Compatible Active Needle

机译:光控MR兼容主动针的设计

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

An active needle is proposed for the development of magnetic resonance imaging (MRI)-guided percutaneous procedures. The needle uses a low-transition-temperature shape memory alloy (LT SMA) wire actuator to produce bending in the distal section of the needle. Actuation is achieved with internal optical heating using laser light transported via optical fibers and side coupled to the LT SMA. A prototype, with a size equivalent to a standard 16-gauge biopsy needle, exhibits significant bending, with a tip deflection of more than 14 in air and 5 in hard tissue. A single-ended optical sensor with a gold-coated tip is developed to measure the curvature independently of temperature. The experimental results in tissue phantoms show that human tissue causes fast heat dissipation from the wire actuator; however, the active needle can compensate for typical targeting errors during prostate biopsy.
机译:提出了一种主动针,用于发展磁共振成像(MRI)引导的经皮手术。针头使用低转变温度形状记忆合金(LT SMA)线致动器在针头的远端产生弯曲。通过使用内部光学加热来实现致动,该内部加热使用通过光纤传输并耦合到LT SMA的激光进行。原型的尺寸相当于标准16口活检针的尺寸,显示出明显的弯曲,在空气中的尖端偏转大于14,在硬组织中的偏转大于5。开发了具有镀金尖端的单端光学传感器,以独立于温度测量曲率。组织体模的实验结果表明,人体组织会导致导线执行器快速散热;但是,活动针可以补偿前列腺活检过程中的典型靶向误差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号