The design and image reconstruction aspects of dynamic SPECT (single-photon-emission computed tomography) instrumentation utilizing multiple rings of position-sensitive photomultiplier tubes (PSPMTs) are reported. A diverging cone-beam collimator is attached to each PSPMT to achieve a 22-cm-diameter spherical field of view in a system designed for head imaging. The resolution with the diverging collimator is estimated to be approximately 20% worse than with a parallel hold collimator. Iterative and direct cone-beam reconstruction algorithms have been investigated under different conditions of angular sampling and counting statistics. The ramifications of temporal resolution and signal-to-noise ratio are discussed in relation to dynamic SPECT imaging of labeled cisplatin for chemotherapy.
机译:报告了利用多个位置敏感光电倍增管(PSPMT)环的动态SPECT(单光子发射计算机断层扫描)仪器的设计和图像重建方面。在为头部成像设计的系统中,将发散的锥束准直仪连接到每个PSPMT,以实现直径22厘米的球形视场。估计发散准直器的分辨率比平行保持准直器的分辨率差约20%。研究了在不同角度采样和计数统计条件下的迭代和直接锥束重建算法。讨论了时间分辨率和信噪比对化学标记顺铂的动态SPECT成像的影响。