Design of a pulsed X-ray system for fluorescent lifetime measurements with a timing accuracy of 109 ps

机译:计时精度为109 ps的用于荧光寿命测量的脉冲X射线系统设计

获取原文
获取原文并翻译 | 示例
           

摘要

We describe the design of a table-top pulsed X-ray system for measuring fluorescent lifetime and wavelength spectra of samples in both crystal and powdered form. The novel element of the system is a light-excited X-ray tube with a tungsten anode at +30 kV potential. The S-20 photocathode is excited by a laser diode with a maximum rate of 10 MHz, each pulse having >100 ps fwhm (full-width at half-maximum) and <10/sup 7/ photons. In a collimated 2 mm/spl times/2 mm beam spot 40 mm from the anode we expect <1 X-ray per pulse. A sample is exposed to these X-rays and fluorescent photons are detected by a microchannel PMT with a photoelectron transit time spread of 60 ps fwhm, a sapphire window, and a bialkali photocathode wavelength range 180-600 nm. The combined time spread of a laser diode, the X-ray tube, and a microchannel tube has been measured to be 109 ps fwhm. To measure fluorescent wavelength spectra, a reflection grating monochromator is placed between the sample and the PMT.
机译:我们描述了一种台式脉冲X射线系统的设计,该系统用于测量晶体和粉末状样品的荧光寿命和波长光谱。该系统的新颖元素是带有30 kV电位的钨阳极的光激发X射线管。 S-20光电阴极由激光二极管以最大10 MHz的速率激发,每个脉冲的功率> 100 ps fwhm(半峰全宽)且<10 / sup 7 /个光子。在距离阳极40 mm的准直2 mm / spl倍/ 2 mm束光斑中,我们期望每个脉冲<1 X射线。样品暴露于这些X射线,并通过具有60 ps fwhm的光电子渡越时间散布,蓝宝石窗口和180-600 nm的双碱光电阴极波长的微通道PMT检测荧光光子。激光二极管,X射线管和微通道管的总时间扩展已测量为109 ps fwhm。为了测量荧光波长光谱,将反射光栅单色仪放在样品和PMT之间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号