Design of a pulsed X-ray system for fluorescent lifetime measurements with a timing accuracy of 109 PS

机译:计时精度为109 PS的脉冲X射线系统设计,用于荧光寿命测量

获取原文

摘要

The design of a table-top pulsed X-ray system for measuring fluorescent lifetime and wavelength spectra of samples in both crystal and powdered form is described. The novel element of the system is a light-excited X-ray tube with a tungsten anode at+30 kV potential. The S-20 photocathode is excited by a laser diode with a maximum rate of 10 MHz, each pulse having >100 ps FWHM and <10/sup 7/ photons. In a collimated 2 mm*2 mm beam spot 40 mm from the anode, <1 X-ray per pulse is expected. A sample is exposed to these X-rays and fluorescent photons are detected by a microchannel PMT (photomultiplier tube) with a photoelectron transit time spread of 60 ps FWHM, a sapphire window, and a bialkali photocathode (wavelength range 180-600 nm). The combined time spread of the laser diode, the X-ray tube, and a microchannel tube has been measured to be 109 ps FWHM. To measure wavelength spectra, a reflection grating monochromator is placed between the sample and the PMT.
机译:描述了一种台式脉冲X射线系统的设计,该系统用于测量晶体和粉末状样品的荧光寿命和波长光谱。该系统的新颖元素是具有+30 kV电位的钨阳极的光激发X射线管。 S-20光电阴极由激光二极管以最大10 MHz的速率激发,每个脉冲具有> 100 ps的FWHM和<10 / sup 7 /个光子。在距阳极40 mm的2 mm * 2 mm准直光束中,每个脉冲的X射线小于1。样品暴露于这些X射线下,并通过微通道PMT(光电倍增管)检测荧光光子,其光电子传输时间扩展为60 ps FWHM,蓝宝石窗口和双碱性光电阴极(波长范围180-600 nm)。测得的激光二极管,X射线管和微通道管的总时间扩展为109 ps FWHM。为了测量波长光谱,在样品和PMT之间放置了一个反射光栅单色仪。