首页> 外文期刊>IEEE Transactions on Nuclear Science >Scatter correction techniques in 3D PET: a Monte Carlo evaluation
【24h】

Scatter correction techniques in 3D PET: a Monte Carlo evaluation

机译:3D PET中的散射校正技术:蒙特卡洛评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

In this work, a Monte Carlo software package, PET-EGS, designed to simulate realistic PET clinical studies, was used to assess three different approaches to scatter correction in 3D PET: analytical (gaussian fitting technique), experimental (dual energy window technique) and probabilistic (Monte Carlo technique). Phantom and clinical studies were carried out by 3D PET and simulated by PET-EGS. A clinical study (/sup 18/F-FDG brain study) was simulated assuming PET emission/transmission multiple-volume images as a voxelised source object describing the distribution of both the radioactivity and attenuation coefficients and accounting for out-of-field activity and media. The accuracy of PET-EGS in modelling the physical response of a 3D PET scanner was assessed by statistical comparison between measured and total (scatter+unscatter) simulated distributions (probability for the two distributions to be the same distribution: p<0.95). The accuracy of the scatter models, for each scatter correction technique, was evaluated on sinograms by statistical comparison between the estimated and the simulated scatter distributions (agreement >1 /spl sigma/). The accuracy of scatter correction was evaluated on sinograms by comparison between scatter corrected and simulated unscatter distributions, proving a comparable accuracy of all the considered scatter correction techniques for brainlike distributed sources.
机译:在这项工作中,设计用于模拟现实的PET临床研究的Monte Carlo软件包PET-EGS用于评估3D PET散射校正的三种不同方法:分析(高斯拟合技术),实验(双能量窗技术)和概率(蒙特卡罗技术)。通过3D PET进行幻影和临床研究,并通过PET-EGS进行模拟。模拟了一项临床研究(/ sup 18 / F-FDG脑研究),假设将PET发射/透射多体积图像作为体素化源对象,描述了放射性和衰减系数的分布,并说明了场外活动和媒体。 PET-EGS在3D PET扫描仪的物理响应建模中的准确性是通过对实测和总(散布+散布)模拟分布(两种分布的概率相同,分布为p <0.95)之间的统计比较进行评估的。对于每种散射校正技术,通过对估计的散射分布和模拟的散射分布(协议> 1 / spl sigma /)之间的统计比较,在正弦图上评估了散射模型的准确性。通过比较散点校正和模拟的非散点分布,在正弦图上评估了散点校正的准确性,证明了所有考虑的散点校正技术对于类似脑部分布源的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号