首页> 外文期刊>Nuclear Science, IEEE Transactions on >Conceptual Design, Development and Preliminary Tests of a Compact Neutron Spectrometer for the JET Experiment
【24h】

Conceptual Design, Development and Preliminary Tests of a Compact Neutron Spectrometer for the JET Experiment

机译:用于JET实验的紧凑型中子光谱仪的概念设计,开发和初步测试

获取原文
获取原文并翻译 | 示例

摘要

A Compact Neutron Spectrometer (CNS) has been developed to measure the neutron emission spectra in Joint European Torus (JET) fusion plasma experiments. The spectrometer, based on a liquid scintillation detector (BC501A), is equipped with a Digital Pulse Shape Discrimination (DPSD) acquisition system for neutron (n) and gamma-ray $(gamma)$ separation. The CNS enables recording the n and $gamma$ pulse height spectra (PHS) up to total count rates of ${sim}10^{6}$ s$^{- 1}$ . Energy resolution, after PHS unfolding, will be ${<}2hbox{%}$ for 14 MeV neutrons and ${<}4hbox{%}$ for 2.5 MeV neutrons. The work done by ENEA-Frascati and Physikalisch-Technische Bundesanstalt (PTB) respectively in the assembly and test of DPSD and scintillation detector, along with the first results obtained by the spectrometer in JET plasma experiments are presented. The experience obtained with CNS in JET will contribute to the development of neutron spectrometers suitable for applications in the International Thermonuclear Experimental Reactor (ITER).
机译:已经开发了紧凑型中子能谱仪(CNS)来测量欧洲联合环形(JET)聚变等离子体实验中的中子发射光谱。该光谱仪基于液体闪烁检测器(BC501A),配备有用于中子(n)和伽马射线(γ)分离的数字脉冲形状鉴别(DPSD)采集系统。 CNS能够记录高达总计数率$ {sim} 10 ^ {6} $ s $ ^ {-1} $的n和$ gamma $脉冲高度谱(PHS)。在PHS展开之后,对于14 MeV中子,能量分辨率将为$ {<} 2hbox {%} $,对于2.5 MeV中子将为$ {<} 4hbox {%} $。介绍了ENEA-Frascati和Physikalisch-Technische Bundesanstalt(PTB)在DPSD和闪烁探测器的组装和测试中所做的工作,以及分光计在JET等离子体实验中获得的第一个结果。 CNS在JET中获得的经验将有助于开发适用于国际热核实验反应堆(ITER)的中子能谱仪。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号