A wireless power transfer (WPT) system based on magnetic resonant coupling is applied to a pacemaker for recharge its battery. The primary coil is assumed to be on-body, while the secondary coil is in-body. Three different configurations of the secondary coil are hereby investigated placing it inside the titanium case of the pacemaker, on the top surface of the case, or being part of the top surface case. The operational frequency is fixed to be at a relatively low frequency (20 kHz) in order to allow field penetration through the case and to limit the electric and magnetic field safety and thermal increase issues. For each examined configuration, these aspects are investigated by numerical and experimental techniques. The obtained results demonstrate the feasibility of the proposed solutions highlighting their advantages and disadvantages.
机译:基于磁共振耦合的无线电力传输(WPT)系统被应用于起搏器以为其电池充电。假定初级线圈在体内,而次级线圈在体内。因此,研究了三种不同配置的次级线圈,将其放置在起搏器的钛金属外壳内,该外壳的上表面或作为上表面外壳的一部分。工作频率固定为相对较低的频率(20 kHz),以允许磁场穿透外壳并限制电场和磁场的安全性以及热量增加的问题。对于每种检查的配置,这些方面均通过数值和实验技术进行研究。获得的结果证明了所提出的解决方案的可行性,突出了它们的优点和缺点。