首页> 外文期刊>IEEE Transactions on Medical Imaging >Unified deadtime correction model for PET
【24h】

Unified deadtime correction model for PET

机译:PET的统一停滞时间校正模型

获取原文
获取原文并翻译 | 示例
           

摘要

A model of deadtime for emission and transmission scans in positron emission tomography (PET) scanners with two-dimensional detectors has been developed. The model takes into account coincidence losses due to singles losses and multiple events, as well as mispositioning errors at higher count rates caused by pulse pile-up, within a detector block. The model is applicable to emission distributions and to spatially varying singles distributions seen with a rotating pin transmission source. An automatic procedure to determine the parameters of this model based on decaying emission studies has also been developed. Different singles dead time factors are required for emission and blank distributions due to differences in their energy spectra. The model was tested on emission and pin transmission data taken on the Scanditronix PC2048-15B scanner.
机译:已经开发了具有二维检测器的正电子发射断层扫描(PET)扫描仪中的发射和透射扫描的停滞时间模型。该模型考虑了由于单次丢失和多个事件引起的一致性损失,以及在检测器模块内由于脉冲堆积而导致的较高计数率下的位置错误。该模型适用于发射分布以及使用旋转销传输源看到的空间变化的单峰分布。还开发了一种基于衰减发射研究确定该模型参数的自动程序。由于它们的能谱不同,发射和空白分布需要不同的单点停滞时间因子。该模型在Scanditronix PC2048-15B扫描仪上进行了发射和引脚传输数据测试。