Metal Artifact Reduction for Polychromatic X-ray CT Based on a Beam-Hardening Corrector

机译:基于束硬化校正器的多色X射线CT金属伪影减少

获取原文
获取原文并翻译 | 示例
           

摘要

This paper proposes a new method to correct beam hardening artifacts caused by the presence of metal in polychromatic X-ray computed tomography (CT) without degrading the intact anatomical images. Metal artifacts due to beam-hardening, which are a consequence of X-ray beam polychromaticity, are becoming an increasingly important issue affecting CT scanning as medical implants become more common in a generally aging population. The associated higher-order beam-hardening factors can be corrected via analysis of the mismatch between measured sinogram data and the ideal forward projectors in CT reconstruction by considering the known geometry of high-attenuation objects. Without prior knowledge of the spectrum parameters or energy-dependent attenuation coefficients, the proposed correction allows the background CT image (i.e., the image before its corruption by metal artifacts) to be extracted from the uncorrected CT image. Computer simulations and phantom experiments demonstrate the effectiveness of the proposed method to alleviate beam hardening artifacts.
机译:本文提出了一种新的方法来校正由多色X射线计算机断层扫描(CT)中金属的存在引起的束硬化伪影,而不会破坏完整的解剖图像。 X射线束多色性的结果是,由于束硬化而导致的金属伪影正成为影响CT扫描的越来越重要的问题,因为医疗植入物在普遍老龄化的人群中越来越普遍。可以通过分析已知的高衰减物体的几何形状,通过分析测量的正弦图数据与理想的前向投影仪之间在CT重建中的失配来校正相关的高阶光束硬化因子。在没有光谱参数或与能量有关的衰减系数的先验知识的情况下,提出的校正允许从未校正的CT图像中提取背景CT图像(即,在其被金属伪影破坏之前的图像)。计算机仿真和幻像实验证明了所提出方法减轻光束硬化伪影的有效性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号