首页> 外文期刊>IEEE Transactions on Magnetics >Development of Peak-to-Peak Voltage Detector-Type MI Gradiometer for Magnetocardiography
【24h】

Development of Peak-to-Peak Voltage Detector-Type MI Gradiometer for Magnetocardiography

机译:磁心动图峰峰值电压检测器型MI梯度仪的研制

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We have developed a high-performance magnetoimpedance (MI) gradiometer with a peak-to-peak voltage detector by synchronized switching (shortened: pk-pk VD-type MI gradiometer) for biomagnetic signal measurement in an unshielded environment at room temperature. This newly developed MI gradiometer detects both the positive peak and the negative peak of the induced voltage waveforms in pick-up coils for enhancing sensitivity and reducing noise level. In this paper, we have investigated the sensitivity and noise level of this new MI gradiometer. The noise level we achieved here is lower than 2 pT/Hz1/2in an unshielded environment for the frequency range of 1-100 Hz. Meanwhile, we have also demonstrated the pk-pk VD-type MI gradiometer for magnetocardiography (MCG) measurements in an unshielded environment and at room temperature with simultaneous measurements of electrocardiography. We have achieved the real-time measurements of R peak in MCG wave using pk-pk VD-type MI gradiometer with an amplitude of 100 pT, when the distance between the MI sensor and the chest surface is 10 mm, in sitting position. Moreover, a clear R wave and T wave of MCG are observed via pk-pk VD-type MI gradiometer, in averaging over six and eight cycles. The room temperature measurements of R wave and T wave in MCG waveform by averaging processing over 10 s (10-15 cycles) without any shielding equipment will contribute to the diagnosis of heart disease.
机译:我们已开发出一种具有峰峰值电压检测器的高性能磁阻(MI)梯度仪,该传感器通过同步切换(简称:pk-pk VD型MI梯度仪)进行了同步,可在室温下的非屏蔽环境中测量生物磁信号。这款新开发的MI梯度计可检测拾波线圈中感应电压波形的正峰值和负峰值,以提高灵敏度并降低噪声水平。在本文中,我们研究了这种新型MI梯度仪的灵敏度和噪声水平。我们在这里达到的噪音水平低于2 pT / Hz n 1/2在非屏蔽环境中,频率范围为1-100 Hz。同时,我们还展示了pk-pk VD型MI梯度仪,用于在非屏蔽环境和室温下同时进行心电图测量的心电图(MCG)测量。当MI传感器和胸部表面之间的距离为10 mm(坐姿)时,我们使用pk-pk VD型MI梯度计(振幅为100 pT)实现了MCG波中R峰的实时测量。此外,通过pk-pk VD型MI梯度计观察到了清晰的MCG的R波和T波,平均周期为六个和八个。在没有任何屏蔽设备的情况下,通过平均处理10 s(10-15个周期)来对MCG波形中的R波和T波进行室温测量,将有助于诊断心脏病。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号