Noncontact ECG Recording System With Real Time Capacitance Measurement for Motion Artifact Reduction

机译:具有实时电容测量功能的非接触式ECG记录系统,可减少运动伪影

获取原文
获取原文并翻译 | 示例
           

摘要

A system for noncontact ECG recording is proposed that measures the real time electrode-body capacitance concurrently with the ECG as a reference signal for motion artifact reduction. Simultaneous recordings of these two signals from the human body in the presence of electrode motion artifacts are shown and an adaptive least-mean-squares (LMS) filtering algorithm run on these signals is demonstrated to be able to reduce the severity of certain types of electrode motion artifacts.