首页> 外文期刊>European Spine Journal >The interobserver-validated relevance of intervertebral spacer materials in MRI artifacting
【24h】

The interobserver-validated relevance of intervertebral spacer materials in MRI artifacting

机译:椎间间隔物材料在MRI人工成像中的经观察者间验证的相关性

获取原文
获取原文并翻译 | 示例
           

摘要

Intervertebral spacers for anterior spine fusion are made of different materials, such as titanium, carbon or cobalt-chrome, which can affect the post-fusion MRI scans. Implant-related susceptibility artifacts can decrease the quality of MRI scans, thwarting proper evaluation. This cadaver study aimed to demonstrate the extent that implant-related MRI artifacting affects the post-fusion evaluation of intervertebral spacers. In a cadaveric porcine spine, we evaluated the post-implantation MRI scans of three intervertebral spacers that differed in shape, material, surface qualities and implantation technique. A spacer made of human cortical bone was used as a control. The median sagittal MRI slice was divided into 12 regions of interest (ROI). No significant differences were found on 15 different MRI sequences read independently by an interobserver-validated team of specialists (P>0.05). Artifact-affected image quality was rated on a score of 0-1-2. A maximum score of 24 points (100%) was possible. Turbo spin echo sequences produced the best scores for all spacers and the control. Only the control achieved a score of 100%. The carbon, titanium and cobalt-chrome spacers scored 83.3, 62.5 and 50%, respectively. Our scoring system allowed us to create an implant-related ranking of MRI scan quality in reference to the control that was independent of artifact dimensions. The carbon spacer had the lowest percentage of susceptibility artifacts. Even with turbo spin echo sequences, the susceptibility artifacts produced by the metallic spacers showed a high degree of variability. Despite optimum sequencing, implant design and material are relevant factors in MRI artifacting.
机译:用于前路脊柱融合的椎间间隔物是由不同的材料制成的,例如钛,碳或钴铬合金,它们会影响融合后的MRI扫描。植入物相关的敏感性伪影会降低MRI扫描的质量,从而妨碍正确的评估。这项尸体研究旨在证明与植入物相关的MRI伪影影响椎间间隔物融合后评估的程度。在尸体猪的脊柱中,我们评估了三个椎间间隔物在植入后的MRI扫描,这些间隔物的形状,材料,表面质量和植入技术均不同。由人皮质骨制成的间隔物用作对照。将中位矢状MRI切片分为12个感兴趣区域(ROI)。经观察员验证的专家组独立读取的15个不同MRI序列上没有发现显着差异(P> 0.05)。受伪影影响的图像质量的评分为0-1-2。最高分数为24分(100%)。 Turbo自旋回波序列为所有间隔物和对照产生了最高分。只有对照达到100%的分数。碳,钛和钴铬间隔物的得分分别为83.3、62.5和50%。我们的评分系统使我们能够参照与假体尺寸无关的对照,创建与植入物相关的MRI扫描质量等级。碳隔离物的磁化伪影百分比最低。即使具有涡轮自旋回波序列,由金属间隔物产生的磁化伪影也显示出高度的可变性。尽管序列优化,但植入物的设计和材料仍是MRI伪影的相关因素。

著录项

  • 来源
    《European Spine Journal》 |2007年第2期|179-185|共7页
  • 作者单位

    Department of Orthopedic Surgery University of Goettingen Robert-Koch-Strasse 40 37075 Goettingen Germany;

    Department of Diagnostic Radiology University of Goettingen Robert-Koch-Strasse 40 Goettingen Germany;

    Department of Orthopedic Surgery University of Goettingen Robert-Koch-Strasse 40 37075 Goettingen Germany;

    Department of Diagnostic Radiology University of Goettingen Robert-Koch-Strasse 40 Goettingen Germany;

    Biomaterials Laboratory/Department of Orthopedic Surgery University of Goettingen Robert-Koch-Strasse 40 Goettingen Germany;

    Department of Orthopedic Surgery University of Goettingen Robert-Koch-Strasse 40 37075 Goettingen Germany;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Intervertebral spacers; Implant materials; MRI; Susceptibility artifacts;

    机译:椎间隙垫片;植入材料;MRI;磁化伪影;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号