Consequences of patient position in the radiographic measurement of artificial disc replacement angles

机译:人工椎间盘置换角射线照相测量中患者位置的后果

获取原文
获取原文并翻译 | 示例
           

摘要

Accurate clinical measurement of spinal range of motion (ROM) is essential in the evaluation of artificial disc performance. The effect of patient placement with respect to the X-ray beam source is yet to be reported and may be an influencing factor in radiographic artificial disc angle measurements. This study aims to evaluate how radiographic patient placement influences artificial disc angle measurements. An anatomically accurate synthetic L4–L5 motion segment was instrumented with an artificial disc and two pins. The instrumented motion segment was mounted onto a frame allowing for independent rotation and elevation while holding the artificial disc angle and anatomical position between L4 and L5 fixed. Analyses included descriptive statistics, evaluation of uncertainty, intra- and inter-observer, and a 2-way analysis of variance (ANOVA). The mean angle measurement range at the various positions was 1.26° for the pin, and 2.74° for the artificial disc endplates. The centered patient position had the highest inter- and intra-observer reliability. ANOVA results showed elevation effects to be statistically significant (P = 0.021), and rotational effects to be extremely statistically significant (P < 0.0001) for the pin angles. In terms of the mean artificial disc angle, however, the ANOVA showed a highly statistically significant interaction term (P = 0.002). A significant difference was found in the angle measurements of a fixed artificial disc prosthesis based on a sample of patient radiographic placement positions. Since it is important to assess the success of an artificial disc replacement by evaluating the relatively small ROM present, it is crucial to aim at minimizing the error by placing the patient parallel to the plate with the beam centered not at the mid lumbar spine, but at the level of the arthroplasty, for both flexion and extension views.
机译:脊柱运动范围(ROM)的准确临床测量对于评估人造椎间盘的性能至关重要。相对于X射线束源的患者放置效果尚待报告,并且可能是放射线照相人造椎间盘角度测量的影响因素。这项研究旨在评估放射照相患者的放置方式如何影响人工椎间盘角度测量。用人造椎间盘和两个销钉对解剖学上精确的合成L4–L5运动段进行了测量。仪器的运动段安装在允许独立旋转和抬高的框架上,同时将人造椎间盘的角度和解剖位置固定在L4和L5之间。分析包括描述性统计,不确定性评估,观察者内部和观察者之间以及方差的2向分析(ANOVA)。销钉在各个位置的平均角度测量范围为1.26°,人造盘端板为2.74°。居中的患者位置在观察者之间和观察者内部具有最高的可靠性。方差分析的结果表明,对于销角,仰角影响具有统计学意义(P = 0.021),而旋转效应具有统计学意义(P <0.0001)。然而,就平均人工椎间盘角度而言,方差分析显示出统计学上高度显着的相互作用项(P = 0.002)。根据患者射线照相放置位置的样本,在固定的人工椎间盘假体的角度测量中发现了显着差异。由于通过评估相对较小的ROM来评估人工椎间盘置换术的成功很重要,因此至关重要的是通过使患者平行于钢板并使光束不居中在腰椎中间,从而最大程度地减少错误。在关节成形术层面上,既可以观察屈曲角度,也可以观察伸展角度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号