In patients with significant left ventricular dysfunction and congestive heart failure despite optimal medical therapy, implantation of cardiac resynchronization therapy-defibrillation (CRT-D) devices has been shown to improve symptoms and mortality. In this report, we describe a case of a patient with ischaemic cardiomyopathy who developed incessant ventricular tachycardia (VT) after undergoing an upgrade from an implantable cardioverter defibrillator to a CRT-D device. The patient required multiple anti-arrhythmic agents, removal of the coronary sinus lead, and radiofrequency ablation to control VT. Thus, in rare patients, the CRT devices may potentially cause ‘proarrhythmia’ with serious consequences.
机译:尽管进行了最佳的药物治疗,但在患有严重左心功能不全和充血性心力衰竭的患者中,植入心脏再同步治疗除颤(CRT-D)设备已被证明可以改善症状和死亡率。在本报告中,我们描述了一例缺血性心肌病患者,该患者经历了从植入式心脏复律除颤器升级到CRT-D设备后不断出现的室性心动过速(VT)。该患者需要多种抗心律不齐药物,去除冠状窦导线和射频消融以控制室速。因此,在罕见的患者中,CRT设备可能会导致“心律失常”并带来严重后果。