Pulmonary vein isolation in a dual-chamber pacemaker patient using the pulmonary vein ablation catheter (PVAC) system resulted in perpetual induction of ventricular tachycardia (VT) during radio frequency energy application. Induction of VT was abolished by programming the PVAC-system to a pure bipolar ablation mode. Patients with implanted devices should be closely monitored when using the PVAC system in unipolar modes.
机译:在双腔起搏器患者中使用肺静脉消融导管(PVAC)系统进行的肺静脉隔离导致在射频能量施加期间永久性诱发室性心动过速(VT)。通过将PVAC系统编程为纯双极消融模式,可以消除VT的诱导。在单极模式下使用PVAC系统时,应密切监视植入了设备的患者。