The implantable cardioverter-defibrillator (ICD) is an effective therapeutic device for preventing mortality in patients with cardiac diseases caused by life-threatening arrhythmias. The authors previously proposed an algorithm for the detection of life-threatening arrhythmias using a multiple regression model. To enhance the classification accuracy, we have now incorporated a moving average filter and a multiple detection process into the original detection algorithm. The experimental results showed that the proposed method could attain sufficiently high accuracy for all ventricular fibrillation rhythms to be detected precisely. In addition, the cases of false detection of sinus rhythm or supraventricular tachyarrhythmias, which could cause ICD malfunction, have been reduced.
机译:植入式心脏复律除颤器(ICD)是一种有效的治疗设备,可预防危及生命的心律不齐导致的心脏病患者的死亡。作者先前提出了使用多元回归模型检测威胁生命的心律不齐的算法。为了提高分类精度,我们现在将移动平均滤波器和多重检测过程合并到原始检测算法中。实验结果表明,所提出的方法可以获得足够高的准确度,可以准确地检测出所有的心室纤颤节律。此外,减少了可能导致ICD故障的误诊窦性心律或室上性快速性心律失常的情况。