Hand-Carried Ultrasonography Instrumentation in the Diagnosis of Temporomandibular Joint Dysfunction

机译:手持超声仪器诊断颞下颌关节功能障碍

获取原文
           

摘要

Internal derangement (ID) in the temporomandibular joint (TMJ) is defined as a mechanical problem of the joint that interferes with its function. It is attributed to an abnormal interaction among the articular disc, condyle, and joint eminence. The aim of this study is to evaluate diagnostic efficacy of non-invasive hand-carried ultrasonography instrumentation (US) to provide high-level images for a correct diagnosis of ID. Twenty-eight ID patients, 15 female and 13 males, were examined both clinically and by MRI images in order to achieve a diagnosis of ID (using Helkimo index). Then, they were submitted to US examination with a 12 MHz transducer by using hand-carried instrumentation by a clinician that was blind to their diagnosis and clinical data. TMJ US examination was performed with the mouth closed and mouth open, with proper technique. Each position was then evaluated with two different orientations of the transducer. US showed acceptable results in identifying bone structures. Lower values of diagnostic efficacy were obtained for disc position during joint movements with respect to MRI images. MRI still represents the gold standard for the identification of joint structures. If not corroborated by clinical and anamnestic data, the diagnostic efficacy of US in identifying the position of the disc during opening and closing jaw movements appears limited than compared to MRI.
机译:颞下颌关节(TMJ)中的内部紊乱(ID)被定义为干扰其功能的接头的机械问题。它归因于关节盘,髁和关节物质之间的异常相互作用。本研究的目的是评估非侵入性手持超声仪器(美国)的诊断效果,以提供高级别的图像,以正确诊断ID。在临床上和MRI图像中检查28名患者,15名女性和13名男性,以实现ID诊断(使用Helkimo指数)。然后,通过使用临床医生对其诊断和临床数据视而不见,通过使用手工携带的仪器提交给我们使用12 MHz传感器进行检查。 TMJ US检查通过闭嘴和嘴巴开放,具有适当的技术。然后用换能器的两种不同取向评估每个位置。美国展示了识别骨骼结构的可接受结果。在相对于MRI图像的关节运动期间,获得较低的诊断效能值。 MRI仍然代表了鉴定联合结构的黄金标准。如果没有通过临床和anamnestic数据证实,我们在打开和关闭钳口运动期间识别盘的位置的诊断功效出现在与MRI相比的限制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号