首页> 外文期刊>Cureus. >A Low-Cost Method to Assess the Performance of Surface Guidance Imaging Systems at Non-Zero Couch Angles
【24h】

A Low-Cost Method to Assess the Performance of Surface Guidance Imaging Systems at Non-Zero Couch Angles

机译:一种低成本的方法,以评估非零沙发角度的表面引导成像系统的性能

获取原文
           

摘要

A procedure is presented to assess performance at non-zero couch angles and perform routine quality assurance (QA) on surface-guided radiotherapy (SGRT) imaging systems used for stereotactic radiosurgery (SRS). A low-cost anthropomorphic phantom was used to assess the system under patient-like conditions. The phantom is embedded with a tungsten ball bearing (BB) to facilitate the use of surface imaging (SI) with concurrent megavoltage (MV) imaging to cross-compare and validate SI-reported offsets. Data analysis is done via in-house software that utilized the SGRT system’s log files for automated analysis. This procedure enables users to assess and inter-compare MV-reported offsets with their SGRT system. The analysis provides SGRT system residual error so that users are aware of inherent offsets present in addition to increases in translational offsets due to couch walkout. The procedure was validated with two commercial SGRT systems. The procedure can be used with any surface imaging system and linear accelerator system.
机译:提出了一种在非零沙发角度评估性能,并在用于立体定向放射外科(SRS)的表面引导放射疗法(SGRT)成像系统上进行常规质量保证(QA)。低成本的拟蒽型幻影用于评估患者样条件下的系统。幻影嵌入钨滚珠轴承(BB),以便于使用表面成像(Si)与并发MegVoLTAGE(MV)成像以交叉比较和验证SI报告的偏移。数据分析是通过内部软件进行的,该软件使用SGRT系统的日志文件进行自动分析。此过程使用户能够使用SGRT系统进行评估和与MV报告的偏移。分析提供了SGRT系统残差误差,以便用户知道除了由于沙发罢工引起的翻译偏移量增加之外,还知道存在的固有偏移。该程序用两个商业SGRT系统验证。该程序可与任何表面成像系统和线性加速器系统一起使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号