Un-LINQed: Spontaneous extrusion of newer generation implantable loop recorders

机译:UN-LINQED:自发挤压较新一代可植入循环录像机

获取原文
获取外文期刊封面目录资料

摘要

BACKGROUND:Insertable cardiac monitors (ICMs) are often used for long-term monitoring of cardiac rhythm. The Medtronic's LINQ Reveal ? is a new generation wireless, automated, and patient responsive subcutaneous ECG monitoring device. Despite several advantages to its small size we have noted an unusually high incidence of extrusion at our center.METHODS:& Results: We conducted a retrospective case analysis to review Reveal LINQs implanted at our center. All devices were inserted using the provided insertion tools. Patients with extruded devices were identified and details regarding the site and technique of insertion, incision closure, use of peri-operative antibiotics, and follow-up details were collected. 81 patients underwent 85 Reveal LINQ implants at a tertiary care University Hospital referral center. The most common reason for implant was suspected arrhythmia with or without structural heart disease or unexplained syncope. There were 4 spontaneous extrusions occurring within 7-24 days after insertion with an incidence rate of 4.7%. One extruded device was anchored to subcutaneous tissue, and no pocket/device infections or hematomas were noted.CONCLUSIONS:Device migration and erosion through skin are important potential adverse events for the Reveal LINQ implantable loop recorder. This study reports an unexpectedly high rate of extrusion without infection. The authors suggest that the depth of the incision is the main factor impacting extrusions. Larger studies are recommended, however, and a proposed measure to avoid spontaneous extrusion is the design of a longer manufacturer's blade in order to increase the depth of the incision and insertion.Copyright ? 2020 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. All rights reserved.
机译:背景:可插入的心脏监测器(ICMS)通常用于长期监测心律。 Medtronic的Linq揭示了?是一个新一代无线,自动化和患者响应皮下心电图监控设备。尽管其小尺寸有几个优势,但我们已经注意到我们的中心异常高的挤出发病率。和结果:我们进行了回顾性案例分析,以审查揭示植入我们中心的LINQs。使用提供的插入工具插入所有器件。鉴定了挤出装置的患者,并收集了有关插入,切口闭合,使用抗生素的使用和随访细节的患者的细节。 81名患者介绍85患者在第三大学医院推荐中心揭示LINQ植入物。植入物最常见的原因是疑似或没有结构性心脏病或无法解释的晕厥的心律失常。插入后7-24天内发生了4种自发挤出率为4.7%。一个挤出装置锚固到皮下组织,没有注意到口袋/装置感染或血液瘤。结论:通过皮肤的装置迁移和侵蚀是显露LINQ可植入环录制仪的重要潜在不良事件。本研究报告了意外的高挤出率而不感染。作者表明切口的深度是影响挤出的主要因素。然而,建议使用较大的研究,并提出避免自发挤出的措施是更长的制造商的刀片设计,以增加切口和插入的深度? 2020印度心脏节奏社会。 Elsevier B.V的生产和托管。保留所有权利。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号