首页> 外文期刊>Scientific reports. >Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)
【24h】

Needle-based storage-phosphor detector radiography is superior to a conventional powder-based storage phosphor detector and a high-resolution screen-film system in small patients (budgerigars and mice)

机译:基于针基的储存 - 磷光体探测器射线照相优于传统的粉末储存磷光体检测器和小患者的高分辨率屏幕膜系统(Budgerigars和小鼠)

获取原文
       

摘要

This method comparison study used radiographs of 20 mice and 20 budgerigars to investigate comparability between computed radiography (CR) and high-resolution screen-film systems and study the effects of reduced radiation doses on image quality of digital radiographs of small patients. Exposure settings used with the mammography screen-film system (SF) were taken as baseline settings. A powder-based storage-phosphor system (CRsubP/sub) and a needle-based storage-phosphor system (CRsubN/sub) were used with the same settings (D/100%) and half the detector dose (D/50%). Using a scoring system four reviewers assessed five criteria per species covering soft tissue and bone structures. Results were evaluated for differences between reviewers (interobserver variability), systems and settings (intersystem variability, using visual grading characteristic analysis). Correlations were significant (p?≤?0.05) for interobserver variability in 86.7% of the cases. Correlation coefficients ranged from 0.206 to 0.772. For mice and budgerigars, the CRsubN/sub system was rated as superior to the SF and CRsubP/sub system for most criteria, being significant in two cases each. Comparing the SF and CRsubP/sub system, the conventional method scored higher for all criteria, in one case significantly. For both species and both digital systems, dose reduction to 50% resulted in significantly worse scores for most criteria. In summary, the needle-based storage-phosphor technique proved to be superior compared to the conventional storage-phosphor and mammography screen-film system. Needle-based detector systems are suitable substitutes for high-resolution screen-film systems when performing diagnostic imaging of small patients. Dose reduction to 50% of the corresponding dose needed in high-resolution film-screen systems cannot be recommended.
机译:该方法比较研究20只小鼠和20个芽刀的射线照片,以研究计算的射线照相(CR)和高分辨率屏幕膜系统之间的可比性,并研究减少辐射剂量对小患者数码射线照片的图像质量的影响。与乳房X线摄影屏幕胶卷系统(SF)一起使用的曝光设置作为基线设置。基于粉末的储存 - 磷光体系(Cr p )和针对基于针对储存 - 磷光体系(Cr n )与相同的设置(D / 100% )和一半的探测器剂量(D / 50%)。使用评分系统四次评审员评估了覆盖软组织和骨结构的每种物种的五个标准。评估了审阅者(Interobserver变异性),系统和设置(使用可视化分级特征分析)之间的差异的差异。相关性(p?≤≤0.05)对于interobserver可变异,在86.7%的情况下。相关系数范围为0.206至0.772。对于小鼠和Budgerigars,Cr n 系统被评为最优于SF和CR P 系统,用于大多数标准,每种情况下都是显着的。比较SF和CR P 系统,传统方法对所有标准进行更高,在一种情况下显着。对于两种物种和数字系统,剂量降低到50%导致大多数标准的得分显着更差。总之,与传统的储存 - 磷光体和乳房X线薄膜系统相比,基于针的储存 - 磷光体技术被证明是优越的。基于针的检测器系统是在进行小患者诊断成像时的高分辨率屏幕膜系统的合适替代品。不能推荐降低高分辨率薄膜系统所需剂量的50%的剂量。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号