Testing and monitoring anabolic androgenic steroids in biological fluids is a key activity in anti-doping practices. In this study, a novel approach is proposed, based on dried urine microsampling through two different workflows: dried urine spots (DUS) and volumetric absorptive microsampling (VAMS). Both techniques can overcome some common drawbacks of urine sampling, such as analyte instability and storage and transportation problems. Using an original, validated liquid chromatography–tandem mass spectrometry (LC-MS/MS) method, exogenous and endogenous unconjugated steroids were analysed. Despite the limitations of microsampling volume, good sensitivity was obtained (limit of quantitation ≤1.5 ng/mL for all analytes), with satisfactory precision (relative standard deviation 70.3%). Both microsampling platforms provide reliable results, in good agreement with those obtained from urine.
机译:在生物流体中的测试和监测合成代谢雄激素类固醇是防掺杂实践中的关键活性。在该研究中,提出了一种基于干燥的尿液微缩穿过两种不同的工作流程的新方法:干燥尿料斑(DUS)和体积吸收微采样(VAMS)。这两种技术都可以克服尿液采样的一些常见缺点,例如分析物不稳定性和储存和运输问题。使用原始的验证的液相色谱 - 串联质谱(LC-MS / MS)方法,分析外源性和内源性的未缀合的类固醇。尽管微小素体积的局限性,获得了良好的灵敏度(所有分析物定量≤1.5ng/ ml),具有令人满意的精度(相对标准偏差70.3%)。两个微内采样平台都提供可靠的结果,与尿液中获得的结果良好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号