首页> 外文期刊>Journal of arrhythmia. >Utility of directional high‐density mapping catheter (AdvisorTM HD Grid) in complex scar‐related atrial tachycardia
【24h】

Utility of directional high‐density mapping catheter (AdvisorTM HD Grid) in complex scar‐related atrial tachycardia

机译:定向高密度映射导管(AdvistormM HD网格)在复杂的瘢痕相关的心房心动过速下的效用

获取原文
           

摘要

Mapping of scar‐related atrial tachycardias (AT) can be challenging even with the use of high‐density (HD) mapping catheter. AdvisorsupTM/sup HD Grid is the only directional HD mapping catheter which not only identify local electrical signal but more importantly capture the direction of wave front propagation especially in low voltage zone. Accordingly, we present a case of complex scar‐related AT with the use of AdvisorsupTM/sup HD Grid which showed clear fractionated signal at isthmus area as compare to the absence of signal on ablation catheter at the same area despite adequate contact force. Ablation at this area terminated the tachycardia.
机译:即使使用高密度(HD)映射导管,疤痕相关的心房心动过速(AT)的映射也可能具有挑战性。顾问 tm 高清网格是唯一的定向HD映射导管,不仅识别本地电信号,而且更重要地捕获尤其是在低压区中的波前传播方向。因此,我们在使用顾问 Tm HD网格上呈现了复杂的瘢痕相关的情况,该HD网格在峡部米地区显示出清晰的分馏信号,如尽管存在于同一区域的消融导管上的信号。足够的接触力。在该地区的消融终止了心动过速。