177Lutetium SPECT/CT: Evaluation of collimator, photopeak and scatter correction

机译:177升性SPECT / CT:准直器,PhotoPak和散射校正的评估

获取原文
       

摘要

Purpose The goal of this study was to find the optimal combination of collimator, photopeak and scatter correction for 177Lutetium (177Lu) SPECT/CT imaging. Methods Three experiments [sphere‐to‐background ratios (SBR) 50:1, 10:1, and 2:1] were performed with the NEMA Image Quality phantom filled with 177Lu‐trichloride. SPECT/CT acquisitions were performed with the medium‐energy low‐penetration (MELP) collimator and 99m?5), whereas the MELP collimator showed a higher CNR for the 2:1 ratio. The RCmean was higher with the MELP collimator, most explicit after the 208‐keV AC/SC reconstruction for all acquisitions. The line profiles showed a better spatial resolution for the MELP collimator and the 208‐keV AC/SC reconstructions. Conclusion 177Lu SPECT/CT image quality and quantification was most optimal when acquired with the MELP collimator and reconstructed using the 208‐keV photopeak, with AC and SC.
机译:目的本研究的目标是找到177升(177LU)SPECT / CT成像的准直器,PhotoPak和散射校正的最佳组合。方法采用填充177万三氯化物的NEMA图像质量幻像进行三个实验[球形 - 背景比率(SBR)50:1,10:1和2:1]。通过中能量低渗透(MELP)准直器和99m?5)进行SPECT / CT采集,而MELP准直器显示出较高的2:1的比例。 MELP准直器的RCMean更高,最明确于所有收购的208 keV AC / SC重建后。线轮廓显示MELP准直器和208-KeV AC / SC重建的更好的空间分辨率。结论当用MELP准直器获得并使用208-KeV Photopak进行重建时,使用MELP准直,使用AC和SC,177Lu Spect / CT图像质量和定量最佳。